Skip to main content
Pump programming

How to I connect my Guardian™ transmitter to my MiniMed™ 700 series pump?

To pair the pump and transmitter:

  • Attach the transmitter to the charger. Fully charge the transmitter. Keep the transmitter attached to the charger.
    Note: Both lights on the charger are off when the transmitter is fully charged.
  • From the Home screen, press the Select button, and then select Paired Devices.
  • Place the transmitter (still attached to the charger) next to the pump.
  • Select Pair New Device. The Searching... screen appears.
    Note: The search process can take up to 20 seconds
  • Remove the transmitter from the charger. The transmitter light flashes 10 times and turns off. The Select Device screen appears with a list of available devices.
  • Select the CGM device that matches the serial number indicated on the back of the transmitter.
  • If the transmitter serial number on the pump screen matches the serial number on the back of the transmitter, select Confirm.

Watch this video showing how to connect your transmitter: https://www.youtube.com/watch?v=9VguJSv7tOY

Need additional help?

Guides and manuals

Still have some doubts?

You can contact us here

Go to the contact page