Skip to main content
Pump programming

How to I connect my Accu-Chek meter to my MiniMed™ 700 series pump?

Watch this video showing how to connect your BG meter: https://www.youtube.com/watch?v=ALpSEKSwodM

The MiniMed™ 780G insulin pump with smart device connectivity can pair with up to four Accu-Chek Guide Link meters.

To prepare the meter to pair with the pump:

  • Press the OK button on the meter to turn on the meter.
  • Select Settings.
  • Select Wireless.
  • Select Yes if the confirmation screen appears on the meter screen. Or, if the confirmation screen does not appear, select Pairing. The meter serial number appears on the meter screen. The meter is now ready to pair with the pump.

To prepare the pump to pair with the meter:

  • From the Home screen, press the Select button, and then select Paired Devices.
  • Select Pair New Device. The Searching... screen appears. After the pump is done searching, the Select Device screen appears.
  • Select the meter that matches the serial number that displays on the meter screen. If the correct serial number does not appear, select Search Again. If the serial numbers shown on the pump screen and meter screen match, select Confirm

If the connection is successful, a Pairing successful message appears on the pump. A Paired with pump message with the serial number of the pump appears on the meter screen.

Watch this video showing how to connect your BG meter: https://www.youtube.com/watch?v=ALpSEKSwodM

Need additional help?

Guides and manuals

Still have some doubts?

You can contact us here

Go to the contact page