What do I do if I receive the Insulin Flow blocked. Fill Tubing stopped. Remove reservoir and select Rewind to restart alert on my pump?

Explanation:

The pump has detected that the insulin flow is blocked while filling the tubing. Possible connection issue between the tubing and reservoir.

Next steps:

Remove the reservoir and select Rewind to restart the fill tubing process.

  • Disconnect the tubing from the reservoir.
  • Confirm that the tubing is not crimped or bent.
  • Continue to follow the steps displayed on the pump using the same infusion set and reservoir.
  • If this alarm occurs again, replace the infusion set